WARNING
1. Intended product use
This attachment should only be used for microscopic observation. Do not use it for any other purpose.
2. Do not disassemble
Disassembly may cause malfunction and/or electrical shock, and will lead to the forfeiture of all claims
against warranty. Do not disassemble any part other than those described in this manual. If you
experience any problem with the microscope, notify your nearest Nikon representative.
3. Input voltage and frequency
• First make sure that the power supply is the correct
model correpoding to your local voltage, either
100/110/120 V or 220/230/240 V, which is shown
below the serial No. on the power supply rear panel.
If it differs, do not turn on the power but contact your
nearest Nikon representative immediately.
• Before turning on the power, look at the input voltage
display and the frequency display located on the power
supply rear panel to check that the values displayed
match the local power to be used. If they differ, be sure
to change the settings before turning on the power.
(P.24) Use of the unit under the wrong power may result
in over-heating due to over-current, fire, electrical
shock or possible damage to the equipment.
<Setting at Shipment>
Units using 100, 110 or 120 V local power: 120 V, 50 Hz
Units using 220, 230, or 240 V local power: 240 V, 50 Hz
4. Use only the specified mercury lamp, lamphouse, fuses, and power cord
Use only the specified mercury lamp, lamphouse, fuses, and power cord.
Use of components other than specified may result in damage to the equipment, fire, or electrical shock.
To prevent an accident of electrical shock, handle carefully the power cord and lamp cord not to subject to
strong shock.
For details on handling lamps, be sure to read the instruction manual supplied with the lamp and follow all
directions contained therein.
• Specified mercury lamp:
• Specified lamphouse:
• Specified fuses:
• Specified power cord:
Safety Precautions
50-W AC mercury lamp HBO50W/AC L1 (manufactured by OSRAM)
C-SHG HG light source 50W
For units using 100/110/120 V local power:
Two 6.3A time-lag, high-breaking type 5 x 20 miniature fuses
For units using 220/230/240 V local power:
Two 4A time-lag, high-breaking type 5 x 20 miniature fuses
For 100/110/120 V AC local power:
• 3-conductor grounding type detachable cordset approved by the
UL
• When using an extension cord, always use one that includes a PE
wire (protective earth wire)
For 220/230/240 V AC local power:
• 3-conductor cordset meeting EU/EN standards
• Use an outlet with protective ground as this is a Protection Class 1
device
E-4
Power supply model
Power Supply
Rear Panel
Frequency Display
Input voltage display