Download Print this page

Endress+Hauser HART iTEMP TMT182B Operating Instructions Manual page 9

Temperature transmitter

Advertisement

iTEMP TMT182B
Endress+Hauser
4
Mounting
4.1
Mounting requirements
4.1.1
Dimensions
For the device dimensions, see the ' T echnical data' →  46.
4.1.2
Mounting location
In the terminal head, flat face, as per DIN EN 50446, direct mounting on insert with cable
entry (middle hole 7 mm).
Make sure there is enough space in the terminal head!
It is also possible to mount the head transmitter on a DIN rail as per IEC 60715 using
the DIN rail clip →  33 accessory.
For information about the conditions (such as the ambient temperature, degree of
protection, climate class etc.) that must be present at the installation point so that the
device can be mounted correctly, see the ' T echnical data' section →  46.
When using in hazardous areas, the limit values of the certificates and approvals must be
observed (see Ex Safety Instructions).
4.2
Mounting the device
A Phillips head screwdriver is required to mount the head transmitter:
• Maximum torque for securing screws = 1 Nm (¾ foot-pound), screwdriver: Pozidriv Z2
• Maximum torque for screw terminals = 0.35 Nm (¼ foot-pound), screwdriver: Pozidriv
Z1
A
8
9
 1
Head transmitter mounting
A
Mounting in a terminal head (terminal head flat face as per DIN 43729)
1
Terminal head
2
Snap rings
3
Insert
B
7
6
5
1
2
4
3
2
1
Mounting
3 4
5
A0046845
9

Advertisement

loading