Download Print this page

Endress+Hauser HART iTEMP TMT182B Operating Instructions Manual page 25

Temperature transmitter

Advertisement

iTEMP TMT182B
Endress+Hauser
8
Commissioning
8.1
Post-installation check
Before commissioning the measuring point make sure that all final checks have been
carried out:
• "Post-installation check" checklist →  11
• "Post-connection check" checklist→  14
8.2
Switching on the transmitter
Once you have completed the post-connection checks, switch on the supply voltage. The
transmitter performs a number of internal test functions after power-up.
The device works after approx. 7 seconds. Normal measuring mode commences as soon as
the switch-on procedure is completed.
8.3
Configuring the measuring device
Wizards
The starting point for device wizards is in the Guidance menu. Wizards not only query
individual parameters but also guide the user through the configuration and/or
verification of entire sets of parameters with step-by-step instructions, including
questions, that are comprehensible for the user. The "Start" button can be disabled for
wizards that require specific access authorization (lock symbol appears on the screen).
The following five operating elements are supported for navigation in the wizards:
• Start
Only on the initial page: start the wizard and go to the first section
• Next
Go to the next page of the wizard. Is not enabled until parameters are entered or
confirmed.
• Back
Return to the previous page
• Cancel
If Cancel is selected, the status before the wizard was started is restored
• Finish
Closes the wizard and possibility of making additional parameter settings on the device.
Only enabled on the final page.
8.3.1
Commissioning wizard
Commissioning is the first step towards using the device for the designated application.
The Commissioning wizard contains an introductory page (with the "Start" operating
element) and a short description of the content. The wizard consists of several sections in
which the user is guided step-by-step through the commissioning of the device.
"Device management" is the first section that appears when the user runs the wizard, and
contains the following parameters. Its main purpose is to provide information about the
device:
Navigation
Guidance → Commissioning → Start
Commissioning
25