Quality Measurement - Siemens Healthineers MAGNETOM Sempra Operator's Manual

Coils
Table of Contents

Advertisement

4
Coils
Performing the measurement
70
1
Align the laser light localizer with the center marking of the coil.
2
Start the measurement.
4.13.3

Quality measurement

The Flex Coil Interface is connected.
The coil is connected to the Flex Coil Interface.
1
Position phantom holder 26 on the Spine 18.
2
Position the Flex coil to be measured with the iPat cushion on the
phantom holder.
3
Attach the coil around the 1900 ml plastic bottle using belts.
Quality measurement Flex Small
Quality measurement Flex Large
4
Align the laser light localizer with the center marking of the coil.
5
Start the quality measurement.
assurance (software))
(
Page 33 Performing quality
Coils | Operator Manual
Print No. M4-08006G.623.01.02.02