Positioning the coil
Positioning the patient
MAGNETOM Sempra
Print No. M4-08006G.623.01.02.02

Use

4.6.2
All RF coils and cushions are removed from the patient table.
1
Position the coil on the patient table with the guide knobs in the
recesses of the T-shaped groove.
2
Connect the coil plug to coil socket 1.
3
Place the positioning aids best suited to the current situation.
4
For patients with small breasts, place a positioning aid into the left
and into the right coil elements.
5
Use the large compression attachment, if necessary.
1
Position the patient in the prone position with the breasts in the
coil elements.
2
Use suitable positioning aids.
3
Set the compression level using the compression control (light,
standard, hard).
The setting is applied to both compression mechanisms.
4
Stabilize each breast with the respective rotary knob.