Siemens Healthineers MAGNETOM Sempra Operator's Manual page 50

Coils
Table of Contents

Advertisement

4
Coils
Performing the measurement
50
1
Align the laser light localizer with the center of the region of
interest.
2
Select the two (up to four, depending on the coil configuration)
corresponding Spine 18 clusters at the syngo Acquisition
Workplace.
3
Start the measurement.
Quality measurement
4.5.3
The quality measurement must be performed in the Head First
orientation. In addition, the measurement must also be performed
with the coil rotated by 180°.
The Spine 18 is positioned on the patient table.
1
Position phantom holder 40 on the table.
2
Place the phantom holder 27 around the 5300 ml plastic bottle.
3
Place the Body 6 around the plastic bottle and secure the coil on
the bottle using the straps.
4
Position the coil with the plastic bottle on the phantom holder 40
so that the coil cable points toward the groove on the side of the
phantom holder.
5
Ensure that the coil cable is routed through the groove on the side
of the phantom holder.
6
Connect the coil plug to a coil socket 1 or 2.
7
Align the laser light localizer with the center of the coil.
8
Start the quality measurement.
assurance (software))
(
Page 33 Performing quality
Coils | Operator Manual
Print No. M4-08006G.623.01.02.02