4
Coils
Positioning the patient
Performing the measurement
56
The Spine 18 is positioned on the patient table.
The head-heel insert and the back-of-the-head support are
positioned on the patient table.
1
Position the base plate at the head end of the patient table. Ensure
that the base plate is seated correctly in the recess.
2
Slide the shoulder coil cup onto the guide rail for the shoulder to
be examined.
3
Slide the shoulder coil cup in the desired position (depending on
the patient´s size).
4
Connect the coil plug of the shoulder coil cup to the coil socket
(left shoulder: 2, right shoulder: 1).
5
Position the corresponding base plate cushion, depending on the
position of the coil cup.
6
Position the shoulder cup cushion in the coil.
7
Place the positioning aids suitable for the current situation.
For example, place the lordosis cushion to bridge the Spine 18 and
the base plate.
1
Position the patient in the supine position with the head to the
magnet.
2
Align the shoulder coil cup on the guide rail so that it adapts to the
shape of the patient's shoulder.
3
Ensure the most isocentric position for the shoulder to be
examined to obtain optimal image quality.
Make sure that the distance between the coil cable and the
patient's head is sufficient.
1
Align the laser light localizer with the center marking of the coil.
2
Start the measurement.
Coils | Operator Manual
Print No. M4-08006G.623.01.02.02