Siemens Healthineers MAGNETOM Sempra Operator's Manual page 24

Coils
Table of Contents

Advertisement

3
General information
Combining coils
Tim Coil Interface
24
9
To remove the coil plug: Grasp the plug housing and lift it up
vertically.
As the Spine 18 is a standard coil for the basic configuration of the
MR system with respect to coil channels, this coil is often mentioned
in the various coil chapters. Some coils are available with different
channel setups, for example, Body 6.
Always observe the message on the Dot display. If a coil code
error is displayed, the connected coil is not released for use on
your system!
If no coil code error is displayed, you can use the coil on this system.
But the number of coil channels used for the measurement is always
limited by the channel setup of the MR system.
For example, on MR systems with a maximum number of 16
channels, you can use a few coil combinations despite the fact that
the combined channels exceed the maximum of 16.
Some coils are valid for several MR systems. Use the Tim Coil Interface
to connect these coils to the MR systems with the new patient table.
(1) Socket cover for coil socket
(2) Release for coils
(3) Interface coil plug
(4) Release for Tim Coil Interface
1
Connect the Tim Coil Interface to the patient table.
Coils | Operator Manual
Print No. M4-08006G.623.01.02.02