Quality measurement

4.4.3
For the Spine 18, the quality measurement is performed separately
for each coil element (SP1 - SP6). The plastic bottle must be
positioned in the center of the respective coil element for each
individual measurement.
1
Position the phantom holder 25 on the Spine 18 in the center of
the corresponding coil element.
2
Align the laser light localizer with the center marking of the
phantom holder.
3
Position the 5300 ml plastic bottle on the phantom holder.
4
Start the quality measurement.
assurance (software))
5
Perform the quality measurement for the coil elements SP1-SP6 in
this way.
4.5

Body 6

Description

4.5.1
The Body 6 enables examinations of the following body regions:
Upper body and abdomen
Pelvis
Depending on the requirements, the coil may be used in the Head
First or Feet First orientation. Several Body 6 coils can be used for
complete upper body examinations.
Coils
(
Page 33 Performing quality
4
47