Siemens Healthineers MAGNETOM Sempra Operator's Manual page 34

Coils
Table of Contents

Advertisement

3
General information
Registering coils
Configuring the quality
measurement
Starting the quality
measurement
34
2
Go to the selection list and select the coil for the quality
measurement.
It may be that some coils are not available in the list. Coils of these
coil types must be registered before starting the quality assurance.
Coil is connected to the patient table.
1
In the status bar, click the Update button to add the type of the
connected coil to the list.
2
Wait at least one minute until the update process is done.
The list is updated automatically.
During the quality measurement, the signal-to-noise ratio and the
signal uniformity of the individual coils are checked.
The coil is selected.
Activate the control box Coil Check.
The quality measurement is configured.
1
Click the Go button to the lower right in the dialog window Quality
Assurance.
The dialog window Workflow Popup opens, including information
for the quality measurement of the coil selected.
2
Follow the instructions for positioning the coil, the phantom, as
well as the phantom holder.
3
Connect the coil to be measured.
4
Click the OK button in the Workflow Popup dialog window.
The dialog window Workflow Popup closes and the quality
measurement starts.
During the measurement, messages regarding the measurement
status are displayed.
Message
ToDo
Running
Status of quality measurement
Measurement still waiting to be performed
Running
Print No. M4-08006G.623.01.02.02
Coils | Operator Manual