Wire The Ma/Sil Output; Wire The Ma/Hart Multidrop Installation - Emerson Micro Motion 5700 Installation Manual

Transmitters with intrinsically safe outputs
Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

Wiring the channels
October 2022
Figure 4-8: Channel B and C mA Output wiring
A. Hazardous area
B. Non-hazardous area
C. mA Output
D. 250–600 Ω resistance
E. 24V nominal
F. Input device
4.9

Wire the mA/SIL output

Refer to the Coriolis Flow Meter with Micro Motion 5700 Transmitters Safety Manual for Safety
Instrumented Systems (SIS) for additional information regarding installation and
commissioning that complies with SIS requirements.
4.10

Wire the mA/HART multidrop installation

Wire the mA/HART multidrop output in explosion-proof, intrinsically safe, or
nonhazardous installations.
WARNING
Meter installation and wiring should be performed only by suitably-trained personnel
using the appropriate government and corporate safety standards.
Procedure
See the following figure for information on wiring an mA/HART multidrop installation.
32
Micro Motion 5700 transmitters with Intrinsically Safe Outputs
Installation Manual
MMI-20039473