Electromagnetic Compatibility; Waste Disposal - Macherey-Nagel URYXXON Relax User Manual

Hide thumbs Also See for URYXXON Relax:
Table of Contents

Advertisement

Available languages

Available languages

15.5 Electromagnetic Compatibility

• Target environment of device use: Application in the commercial laboratory environment.
• This device has been developed and tested in accordance with CISPR 11 Class A. It can
cause radio interference in a domestic environment. In this case, you may need to take
action to correct the interference.
• Use of this equipment in a dry environment, especially if synthetic materials (synthetic fiber
clothing, carpets, etc.) are present, can cause destructive static discharges that can cause
erroneous results.
• Do not use this device near sources of strong electromagnetic radiation, as they may
interfere with proper operation.
• This in vitro diagnostic device is in accordance with the applicable parts of the requirements
for emission and immunity described in the IEC 61326 series of standards.
• Before use, the electromagnetic environment must be assessed.
EMC Class A according CISPR 11:
The device can be used in all institutions excluding residential areas and areas which are
directly connected to the public power supply, if the puplic power supply supplies buildings for
residential purpose.

15.6 Waste disposal

Disposal according to EU Directive 2012/19/EU. In accordance with the EU Directive
2012/19/EU MACHEREY-NAGEL takes back the old device and disposes it free of
charge.
Disposal using public waste disposal facilities is not permitted. In case of disposal, please
contact your MACHEREY-NAGEL representative.
NOTICE
Manual URYXXON
®
Relax EN, V 6.03 / 08.22
109