Reported Value - (Original Threshold) - Modification; How To Update The Instrument; Control Strip - Macherey-Nagel URYXXON Relax User Manual

Hide thumbs Also See for URYXXON Relax:
Table of Contents

Advertisement

Available languages

Available languages

11.3.3 Reported value – (original threshold) – modification
Press
to select the value that you would like to modify. Three informations are in each
line:
The information is always defined "X(Y)Z", e.g. "NEG (650) 0"
• X: Measurement value "NEG"
• Y: Specific border "650"
• Z: Change "0"
Press the buttons
and
to increase or decrease the threshold value. Press
to
+
0
return to the original threshold value.
Confirm your modifications by pressing
. For the parameter Microalbumin the reported
values for Kreatinin and Albumin have to be changed separately.

11.4 How to update the instrument

Select "Program update" from the service menu. The URYXXON
Relax now expects the
®
upload of a new firmware. Follow the instructions that come with the update-file to finalize the
update process.

11.5 Control strip

The check mode for control strips allows you to check the correct functioning of the instrument
using special color control strips. You can order these control strips from MACHEREY-NAGEL
(REF 930988).
Choose the round button Control strip in the check mode measurement menu (Display 25).
The instrument asks for insertion of the colored control strip (Display 33). The measurement of
the control strip starts automatically as soon as the strip is placed on the sled, if the autostart is
activated. Once the measurement has been completed, the measured values are immediately
displayed and / or printed. The result of the measurement is not stored. The values are so
called remission values for the pre-colored test pads on the control strip. These values have
to be compared to a set of standard values from the control strip package instructions. For
additional information on the control strips or in case the values do not fit the should values in
the control sheet, please consult the package insert, or contact MACHEREY-NAGEL directly.
ATTENTION: Do not touch the test fields of the control strip with your fingers.
MEASUREMENT
Insert
Control strip
Display 33: Insert control strip
Manual URYXXON
Relax EN, V 6.03 / 08.22
®
97

Advertisement

Table of Contents
loading

Table of Contents