Chapter 12
Technical Specifications

Front Panel Keys

12-6
Patient Data Range, Resolution, Units & Accuracy (Continued)
Patient Data
Display Range
I:E
9.9:1 to 1:99
Peak Inspiration
-20.0 to 130
Pressure
End Inspiration
-20.0 to 130
Pressure (Pi End)
Mean Airway
-20.0 to 120
Pressure
Delivered O
%
0.0 to 110
2
Total Respiratory Rate
0.0 to 150
Spontaneous
0.0 to 150
Respiratory Rate
End Exhalation
-20.0 to 99.9
Pressure (Pe End)
Patient Leak (Pt Leak)
0.0 to 140
Percent Patient Trigger
0.0 to 99.9
(%Pt Trigger)
Table 12-10: Patient Data Range, Resolution, Units & Accuracy (Sheet 2 of 2)
a. Due to the variable nature of patient leaks, this parameter is an estimate only.
Alarm Silence
Alarm Reset
100% O
2
Manual breath
Esprit® Ventilator Operator's Manual © Respironics, Inc.
Units
Resolution
none
0.1 for 9.9:1-1:9.9;
1 for 1:10-1:99
cmH
O
1 for 100-130;
2
(hPa)
0.1 for -20.0 to 99.9
cmH
O
1 for 100-130;
2
(hPa)
0.1 for -20.0 to 99.9
cmH
O
1 for 100-120;
2
(hPa)
0.1 for -20.0 to 99.9
--
1 for 100-110;
0.1 for -20.0 to 99.9
Bpm
1 for 10-150;
0.1 for 0.0 to 9.9
Bpm
1 for 10-150;
0.1 for 0.0 to 9.9
cmH
O
0.1
2
(hPa)
LPM
0.1 for 0.0-99.9
1 for 100-140
%
0.1 for 0.0-99.9
1 for 100
Front Panel Keys
Screen Lock
Exp. Hold
Options
Accept
Table 12-11: Front Panel Keys
Accuracy
±10%
±10%
±10%
±10%
±3 vol%
±10%
±10%
±10%
Not
a
specified
±10%
REF 580-1000-01 Y