Chapter 8
Operating Instructions

Patient Data Screen

8-26
You can view the patient data screen (Figure 8-30) by pressing the PATIENT
DATA button in the top bar. The format of this screen is the same in VCV, PCV,
and NPPV. Table 8-3: "Patient Data Definitions: Range, Units & Resolution"
summarizes the definitions, display ranges, and resolution for patient data
parameters.
Figure 8-30: Patient Data Screen
Patient Data Definitions: Range, Units & Resolution
Parameter
Description
PIP
Peak Inspiratory Pressure
The maximum airway pressure
during inspiration.
MAP
Mean Airway Pressure
The average of airway pressure
during one full breath cycle.
Pe End
Pressure at End Expiration
The pressure measured at the end
of expiration.
Pi End
Pressure at End Inspiration
The pressure measured at the end
of inspiration (or at the end of Insp
Hold if it is >0).
Tidal Vol.
Exhaled Tidal Volume
The volume exhaled at each breath.
Compliance compensated (if
enabled).
Spont VE
Spontaneous Minute Volume
The spontaneous ventilation
normalized to one minute.
Compliance compensated (if
enabled).
Table 8-3: Patient Data Definitions: Range, Units & Resolution (Sheet 1 of 2)
Esprit® Ventilator Operator's Manual © Respironics, Inc.
Display Range
Resolution
-20.0 to 130
0.1 for -20.0 to 99.9
cmH
O (hPa)
1 for 100 to 130
2
-20.0 to 120
0.1 for -20.0 to 99.9
cmH
O (hPa)
1 for 100 to 130
2
-20.0 to 99.9
0.1
cmH
O (hPa)
2
-20.0 to 130
0.1 for -20.0 to 99.9
cmH
O (hPa)
1 for 100 to 130
2
0 to 9999 ml
1
(BTPS)
0.00 to 99.9 L
0.01 for 0.00 to 9.99
(BTPS)
0.1 for 10.0 to 99.9
REF 580-1000-01 Y