Pressure Control Ventilation Settings, Ranges & Resolution - Respironics ESPRIT Operator's Manual

Hide thumbs Also See for ESPRIT:
Table of Contents

Advertisement

Chapter 12
Technical Specifications
Pressure Control
Ventilation Settings,
Ranges & Resolution
12-2
Volume Ventilation Settings, Ranges & Resolution (Continued)
Setting
Range
Inspiratory Trigger
(I-trigger)
Pressure Sensitivity
-20 to -0.1 cmH
Flow Sensitivity
0.5 to 20 LPM (Resolution is 0.1 LPM)
Expiratory Trigger
(E-Cycle)
% Peak Flow
10 to 80% of inspiratory peak flow
Rise Time
0.1 to 0.9 seconds (Resolution is 0.1 second)
%O
21% to 100%
2
Insp. Hold
0 to 2.0 seconds (Resolution is 0.1 second)
Flow Waveform
Descending ramp, square
Patient Type
Adult/Pediatric
Table 12-3: Volume Ventilation Settings, Ranges & Resolution (Sheet 2 of 2)
Resolution is one unit unless otherwise noted (cmH
numerically equivalent to hPa).
Pressure Control Ventilation Settings, Ranges & Resolution
Setting
Range
Apnea Rate
1-80 Bpm
Respiratory Rate
1 to 80 Bpm
PCV Pressure
5 to 100 cmH
Inspiratory Time
0.1 to 9.9 seconds (Resolution is 1 second)
PEEP
0 to 35 cmH
PSV Pressure
0 to 100 cmH
Inspiratory Trigger
(I-trigger)
Pressure Sensitivity
-20 to -0.1 cmH2O (hPa) (Resolution is 0.1 cmH2O (hPa))
Flow Sensitivity
0.5 to 20 LPM from base flow of 3 LPM above sensitivity (Resolution is
0.1 LPM)
Expiratory Trigger
(E-Cycle)
% Peak Flow
10 to 80% of inspiratory peak flow
Table 12-4: Pressure Control Ventilation Settings, Ranges & Resolution (Sheet 1 of 2)
Esprit® Ventilator Operator's Manual © Respironics, Inc.
O (hPa) (Resolution is 0.1 cmH
2
O is considered
2
O (hPa) (Relative to PEEP)
2
O (hPa)
2
O (hPa) (Relative to PEEP)
2
O (hPa))
2
REF 580-1000-01 Y