Non-Invasive Positive Pressure Ventilation Settings Report (Npvs - Respironics ESPRIT Operator's Manual

Hide thumbs Also See for ESPRIT:
Table of Contents

Advertisement

Appendix A
RS-232 Communications Protocol
Non-Invasive Positive
Pressure Ventilation
Settings Report
(NPVS)
Appendix A-16
Pressure Control Ventilation Settings Report (Continued)
Description
Example
PCV High Minute
60.0
Volume Alarm Limit
Setting
Stop Code
0x03
Table A-3: Pressure Control Ventilation Settings Report (Sheet 3 of 3)
When the ventilator receives NPVS followed by a carriage return, it will respond
by transmitting the information shown in Table A-4: "Non-Invasive Positive
Pressure Ventilation Settings Report". The ventilator responds to the NPVS
command by returning a string with a variable length. Fields 2 through 4
define the length of the message. The last character transmitted is a stop code
indicating the end of the message. The second field indicates the number of
characters between the start and stop codes. The third field indicates the
number of fields between the start and stop codes. The fourth field is the start
code, 0x02. The last field in the string is the stop code, 0x03.
Non-Invasive Positive Pressure Ventilation Settings Report
Description
Example
Command Name
NPVS
Number of characters
168
between the start and
stop codes
Number of fields
23
between the start and
stop codes
Start Code
0x02
Time of request
13:45
Date
FEB 23 1997
Current Ventilation
VCV
Type
Table A-4: Non-Invasive Positive Pressure Ventilation Settings Report (Sheet 1 of 3)
Esprit® Ventilator Operator's Manual © Respironics, Inc.
Resolution Range
Units
0.01 for
0.00 - 60.0 L
0.00 to
9.99
0.1 for
10.0 to
60.0
N/A
N/A
N/A
Resolution Range
Units
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
VCV
N/A
PCV
NPPV
Comments
ASCII End
Transmission
Character (ETX)
Comments
3 character field
2 character field
ASCII Start
Transmission
Character (STX)
24 hour clock,
hh:mm
12 character field,
MMM DD YYYY
The ventilation
type currently
being used by the
ventilator.
REF 580-1000-01 Y