NICO-Esprit Interface Option

Trended NICO Data

13-104
Information Transfer from the
The Esprit ventilator provides the NICO monitor with the following information.
Information Transferred from the Esprit Ventilator to the NICO Monitor
Data
Description
Breath Type
Mandatory, Assisted, Spontaneous, or Supported
FIO
The percentage of oxygen in the gas delivered to the patient.
2
NOTE:
Table 13-15: Information Transferred from the Esprit Ventilator to the NICO Monitor
The Esprit ventilator captures the data sent from the NICO monitor and stores
it on the Esprit PCMCIA card (PC Card). The Trending screen makes it possible
to display ventilator and monitor information for intervals from 2 to 72 hours.
To access NICO monitor trended data:
1. Touch the Patient Data button at the top of any settings screen (VCV,
PCV or NPPV).
NOTE:
If the Graphics option is not installed, the Graphics button is grayed out
and Trending is only accessible from the Patient Data screen.
Esprit® Ventilator Operator's Manual © Respironics, Inc.
Esprit Ventilator to the NICO Monitor
If the optional oxygen sensor for the Esprit
ventilator is used, this number will be the
measured by the sensor. If the
FIO
2
optional oxygen sensor is not used, this will
be the set
.
FIO
2
REF 580-1000-01 Y