Appendix A
RS-232 Communications Protocol
Appendix A-4
Description
Example
PCV Inspiratory Time
1.00
Setting
PCV PEEP Setting
0
PCV Pressure Support
0
Setting
40PCV I-Trigger Type
PRESS
PCV Pressure I-Trigger
2.0
Setting
PCV Flow I-Trigger
3.0
Setting
PCV E-Cycle Type
%FLOW
PCV %Flow Expiratory
25
Cycle Setting
PCV Rise Time
0.1
PCV Oxygen
21
Concentration Setting
PCV Apnea Rate
12
Setting
PCV Apnea Interval
20
Alarm Limit Setting
PCV High Inspiratory
35
Pressure Alarm Limit
Setting
PCV Low Inspiratory
3
Pressure Alarm Limit
Setting
PCV Low PEEP Alarm
0
Limit Setting
PCV Low Mandatory
0
Tidal Volume Alarm
Limit Setting
PCV Low Spontaneous
0
Tidal Volume Alarm
Limit Setting
PCV High Respiratory
150
Rate Alarm Limit
Setting
Table A-1: Ventilation Report (Sheet 3 of 10)
Esprit® Ventilator Operator's Manual © Respironics, Inc.
Ventilation Report (Continued)
Resolution Range
0.01
0.1 - 9.9
1
0 - 35
1
0 - 100
N/A
PRESS
FLOW
AUTO
0.1
0.1 - 20.0
0.1
0.3 - 20.0
N/A
%FLOW
AUTO
1
10 - 80
0.1
0.1 - 0.9
1
21 - 100
1
1 - 80
1
10 - 60
1
10 - 105
1
3 - 105
1
0 - 35
1
0 - 2500
1
0 - 2500
1
0 - 150
Units
Comments
sec
cmH
O
2
(hPa)
cmH
O
2
(hPa)
N/A
cmH
O
2
(hPa)
LPM
N/A
%
sec
%
BPM
sec
cmH
O
2
(hPa)
cmH
O
2
(hPa)
cmH
O
2
(hPa)
mL
mL
BPM
REF 580-1000-01 Y