Chapter 13
Oxygen Manifold Option

Assembly Instructions

13-24
WARNING:
The Oxygen Manifold is for oxygen use only
WARNING:
Maximum inlet pressure: 90 PSIG
Minimum inlet pressure: 50 PSIG at 200 SLPM
WARNING:
Keep all hoses within the limits of the ventilator footprint to prevent
a tripping or snag hazard.
CAUTION:
Use this kit with the Esprit ventilator only.
CAUTION:
Do not use the oxygen manifold or oxygen hoses to move the ventilator
NOTE:
Compatible with Esprit-Software version 3.1 or higher only
If a Cord Storage Bracket is not installed (refer to Figure 13-4), ensure that
pre-drilled holes (refer to Figure 13-7) are present. If holes are not present,
you will not be able to assemble the O
800-345-6443 and request help from RI Service.
1. Lock the front wheels of the ventilator.
Figure 13-2: Locking Ventilator wheel
2. Attach the oxygen manifold assembly to the ventilator by hand,
threading the adapter fitting onto the oxygen inlet elbow fitting (refer
to Figure 13-3).
Esprit® Ventilator Operator's Manual © Respironics, Inc.
kit. Contact customer service at 1-
2
REF 580-1000-01 Y