5.3.2 ECG Display

If ECG display is set to on in advanced setting menu, only one second is needed to shift to
real-time waveform display mode by pressing left-up key and confirmation key together. Users
can observe the validity of recorded waveform.
Display one lead (channel) ECG waveform, bottom center displays the name of lead (channel) (I,
II, III, aVR, aVF, aVL, V1, V2, V3, V4, V5, V6 or CH1, CH2, CH3 under 3-channle mode).
Press left-up key or right-down key to change ECG lead (channel), press left-up key and
confirmation key together to shift to information display mode.

5.4 Stopping Recording

1. Recording will automatically stop if recording has lasted for 1 day (or 2 days, 3 days set
beforehand);
2. Recording will automatically stop if battery is used up;
3. Recording will automatically stop due to limited SD card capacity;
4. Directly connect with USB cable after removing patient cable;
5. Press confirmation key, left-up key and right-down key together to stop recording.
The above are recommended regular ways to terminate recording.
Directly taking out battery will also terminate recording, while it may cause damages to data. SD
card may have to be scanned and repaired through disk scanning program of operating system,
which shall be avoided.
Take out SD card and insert it into analysis system card reader or connect USB cable with
recorder to prepare for analysis;
Remove leads and electrodes from patients and do necessary cleaning for them.