Battery Loading; Features - EDAN SE-2003 User Manual

Holter system recorder
Table of Contents

Advertisement

SE-2003&SE-2012 Holter System Recorder User Manual

2.5 Battery Loading

SE-2003/SE-2012 is powered by one AAA battery. Large capacity AAA alkaline battery is
recommended. Press
push outwards with force. The cover of battery compartment will open. Load one AAA battery
according to polarity indication inside battery compartment.
NOTE: There is battery type setting in the setting menu of SE-2003/SE-2012: alkaline
battery and Ni-MH battery. The purpose is to give more exact warning message
on battery under-voltage according to discharge performance of different battery
types. Please set the menu according to used battery type.
1. Take out battery if the recorder is not to be used for a long time. Or else damage from
corrosion could result.
2. Do not throw off used or scrapped batteries. Observe operation instructions and local
laws for battery disposal.

2.6 Features

Refined appearance: 68 x 53 x16 mm, 42g
SE-2012 Holter System Recorder is able to create standard 12 leads or 3 channels
complete ECG data;
EDAN's unique multi-channel pacemaker detect circuit effectively prevents wrong
detection of pacemaker signal caused by all kinds of artifacts (such as movement,
polarized voltage and skin impedance) and missing detection caused by software
detecting only. The detection sensitivity can reach 10
128 X 64 graphic LCD plus a 3-key keyboard make it easy to set recording parameters
of the Holter System Recorder. Real time ECG waves display helps to check electrode
placement quality. During recording, you can switch to ECG display window at any time
to master ECG recording situation;
Multi-languages menu, easy and friendly to operate;
Real-time clock, real-time display of year, month and date; the recording time is actual
time, which prevents trouble and poor accuracy caused by recording manually.
E-label: Support either registration in analysis software or entry of patient ID by recorder
keyboard. Basic information (patient ID, name, gender, age) of patients is written into
data package before making records. In this way recorders used by different patients will
mark on battery compartment on the back of recorder by thumb and
WARNING
-4
second.
- 9 -