2.0 - Calibrator Operation

The calibrator keeps the block temperature controlled and allows the
reading of a thermo-element connected to its terminals. It is possible to verify
the thermocouple, thermoresistances, temperature transmitters etc. together
with the block temperature value and temperature set point.
The calibrator has 3 operating modes:
• Manual Mode to select the block temperature straight from the keypad.
• Programmable Mode: 6 different programs with 11 temperature set point
values. The block temperature is selected among the programmed values by
the keys
and
the temperature set points.
• Automatic Mode to calibrate thermo-elements. The thermo-elements
calibration is performed in an automatic way by the calibrator: the schedule
and the calibration results, besides the thermo-element readings are stored in
the calibrator memory.
When powered on (ON/OFF key), the calibrator goes through a self-test
routine and shows the last calibration date and the value of the battery voltage. The
battery voltage is constantly monitored and the low battery warning is provided.
After the self- test is completed, the display shows the starting menu. he initial set
point is 25.00 ºC:
Através das teclas
By means of keys
ENTER.
IN: selection of the calibrator signal input. Choose among mV, Ohms,
thermocouples, thermoresistances, mA, switch or none. More details in item 2.1 - Menu IN.
EXEC: The calibrator enters the manual or programmable operating mode.
CONF: Accesses the calibrator configuration options. More details in item
2.2 - Menu CONF.
CAL: This option accesses the
functions, protected by password. More details in section 6 - Calibration.
COM: Accesses the automatic calibration parameters. It is possible to perform a
calibration without a computer or using it (via the ISOPLAN software). More details in item
2.5 - Automatic Operating Mode.
Page 8
. The Programmable Mode with Timer scans automatically
IN
EXEC
CONF
CAL
, , 
e
, , 
TE -25N / TE-35N / TE-50N
COM
, escolha as opções do menu e tecle ENTER.
and
, choose the menu options and press
TE-25N/TE-35N/TE-50N
calibrator adjust

Advertisement

Table of Contents
loading

This manual is also suitable for:

Te-35nTe-50n

Table of Contents