performed by the calibrator. The keys
the block and thermo-element temperature value and the other displays a message
indicating the calibration point number and its state (OK or FAIL), besides the error value
in %.
The option TAG from menu COM can present a list of up to 4 tags that have been
downloaded or uploaded with the ISOPLAN-5 software. The key ENTER selects the tag to
be calibrated from the tag list. The automatic calibration starts by the option EXEC from
menu COM and the calibration data can be verified in option VERIF. To calibrate
automatically not using ISOPLAN-5, there is a reserved place for one manual tag. The
selection of the manual tag is performed by confirming the option TAGMAN with the key
ENTER.
The option ADDRESS selects the communication address of the TE-25N/TE-
35N/TE-50N calibrator. The communication protocol used is ModBus - RTU, with no parity
and baud rate of 9600. The calibrator communication with the computer can use RS-232 or
RS- 485, for network option, according to the communication interface used. In order to
communicate with the ISOPLAN software, configure ADDRESS to 1.