INPUT HIGH:20.0000 mA
INPUT LOW: 4.0000 mA
SCALE DEC (0-4): 1
SCALE HIGH: 100.0 #
SCALE LOW: 0.0 #
MEM: The TE-25N/TE-35N/TE-50N calibrator allows many special programs and
functions that can be of frequent use. In situations like this, it would be useful to store the
current configuration in memory in order to save time. Up to 8 configurations can be stored
in memory.
Selecting the option MEM, it is possible to store the current configuration (WRITE),
restore a previous stored configuration (RECALL) or erase the 8 configurations from
memory (CLEAR ALL).
Selecting the option WRITE or RECALL will present a new menu with numbers 1 to
8, representing each one of the memory positions. Choose one of the positions and press
ENTER. The writing operation (WRITE) can be made in an already used memory position.
The calibrator asks for the overwriting confirmation with the message OVERWRITE
MEMORY?. The CLEAR ALL operation shows a confirming message ARE YOU SURE?.
In both cases, press ENTER to confirm the operation or C/CE to cancel.
WRITE
CLEARALL
TE -25N / TE-35N / TE-50N
RECALL
Page 13