7.0 - Instructions for Hardware Maintenance

There are no parts or components in the temperature calibrator that can be repaired
by the user. Only the 6Amp fuse, placed within the socket on the rear can be replaced by the
user.
The fuse may blow due to a voltage spike in the mains or a calibrator component fault.
Replace the fuse once. If a second fuse blows again, it is because the fault is not that simple.
Send the calibrator to the factory for repair.

8.0 - Instructions for Fitting the Strap for Transport

The strap for transport is included together with the dry block. It is very useful during
the transport period and its fitting is very simple. Just manually screw the screws on each tip
of the strap in the holes located in the dry block sides. See illustration below.
Take care when using the cup-like insert and the tiny balls of steel. Handle the balls or
the sensors only when they are at ambient temperature. Operate the dry block calibrator in a
proper location to prevent it from falling down or even to prevent the balls at high
temperature from being spilled outside causing body burn or other damages or injuries.