Chapter 5: Replacing Assemblies
To remove and replace the cover, top plate, and bottom plate
To remove and replace the cover, top plate, and bottom plate
Use this procedure to remove and replace the cover, top plate, and bottom
plate. When necessary, refer to other removal procedures. The pictures in this
chapter are representative of the oscilloscope at the time of this printing. Your
unit may look different.
Disconnect the power cable.
1
Disconnect all oscilloscope probes and BNC input cables from the front panel.
2
Disconnect any other cables, such as mouse, keyboard, printer, USB, or LAN
3
cables.
Remove the accessory pouch from the top of the oscilloscope if it is attached.
4
Remove the two M4 screws securing the side handles on both sides of the
5
oscilloscope. When installing, torque to 18 in- lbs.
Figure 5-1
Remove the six M4 screws that secure the rear feet (three in each foot). When
6
installing, torque to 18 in- lbs.
Figure 5-2
82