Chapter 4: Troubleshooting

Acquisition/Backplane Assembly Troubleshooting

Acquisition/Backplane Assembly Troubleshooting
This section describes which board assembly to replace if any of the scope self tests
fail. When the self- test error message file is generated it is sent to the following location:
C:\ProgramData\Agilent\Infiniium\selftest\selftestlog.txt
The error message usually indicates the channel with the error. Channels 1 and 2 are
on the lower acquisition board. Channels 3 and 4 are on the upper acquisition board.
Replace the acquisition assembly that has the error.
If the error message does not indicate a channel, refer to the following table to determine
which assembly to replace.
Test Group and Test Name
Timebase Test Group
Timebase Interpolator Test
ADC Test Group
ADC Register Tests
ADC Voltage Test Points
Acquisition Memory Test Group
Hedwig MBIST
Hedwig Register Tests
Hedwig-DDR2 Addr & Data
Test
Hedwig DDR2 Stress Test
Hedwig DDR2 Interface
Misc. Scope Test Group
Temp Sensor Tests
Acq Flash RAM Tests
70
Error Type
ADC1, ADC2, ADC3, ADC4
ADC5, ADC6, ADC7, ADC8
ADC1, ADC2, ADC3, ADC4
ADC5, ADC6, ADC7, ADC8
MEMCON_1, MEMCON_2, MEMCON_3,
MEMCON_4
MEMCON_5, MEMCON_6, MEMCON_7,
MEMCON_8
Memory Controller 1-4
Memory Controller 5-8
Hedwig1, Hedwig2, Hedwig3, Hedwig4
Hedwig5, Hedwig6,Hedwig7, Hedwig8
Hedwig0, Hedwig1, Hedwig2, Hedwig3, Oak0
Hedwig4, Hedwig5, Hedwig6, Hedwig7, Oak2
Pred1, Pred2, MainFPGA
Test is currently not being run.
Assembly to Replace
Backplane assembly
Lower acquisition assembly
Upper acquisition assembly
Lower acquisition assembly
Upper acquisition assembly
Lower acquisition assembly
Upper acquisition assembly
Lower acquisition assembly
Upper acquisition assembly
Lower acquisition assembly
Upper acquisition assembly
Lower acquisition assembly
Upper acquisition assembly
Backplane assembly