Managing patients and monitoring
1. Insert the Hub into the pouch. Check that the Hub parts indicated with the
exclamation mark are aligned with the pouch as in the graphic.
It is especially important that the speaker is not covered by the pouch to avoid
missing any audible alarms.
2. Instruct the patient that they can wear the Hub pouch when moving around with
the Hub. The Hub pouch can be worn in different ways: neck, shoulder, back etc.
If the pouch or Hub get wet, dry them to ensure correct functioning of the speaker
and touchscreen. Always replace a dirty or broken Hub pouch.

Unassigning the patient from the Hub

Unassigning a patient from a Hub ends the monitoring, deletes the patient data from
this Hub, and clears the sensor(s) and Hub pairing. It also allows the Hub to be used
for another patient. You can do this at either the Hub or the Central Viewer.
1. Unassign the patient using the Hub:
a. Activate the screen and enter the passcode.
b. Tap the patient name to access the patient detailed view.
c. Select End monitoring.
A message appears asking you to confirm that you wish to end monitoring.
d. Press
upper right corner of the message to return to the previous view without
clearing the Hub and patient assignment.
The assignment between the patient and the Hub is cleared. The message Patient
data removed appears. The patient stays admitted to the monitoring system until
discharged from the Central Viewer.
2. Unassign the patient from the Hub using the Central Viewer:
a. Select a patient slot in the Multiple Patient View.
b. From the Single Patient View action menu, select Remove Hub.
The Remove Hub? menu appears.
c. Select Yes, Remove to unassign the patient from the Hub, or select Cancel to
return to the previous screen without clearing the Hub and patient assignment.
After selecting Yes, Remove, the assignment between the patient and the Hub
is cleared and the devices can be used for a different patient after cleaning and
disinfection. The patient stays admitted to the monitoring system until discharged
from the Central Viewer.

Shutting down the Hub

Make sure that the Hub is not needed for monitoring a patient before shutting it down.
The Hub is shut down with the power key (1.) located on the side of the Hub:
70
to unassign the patient from the Hub, or select the X on the
Portrait
Mobile Monitoring Solution
5815148-01