Operation Warnings - GE Portrait Hub01 User Manual

Mobile monitoring solution
Table of Contents

Advertisement

WARNING
WARNING
WARNING
WARNING
WARNING
WARNING

Operation warnings

WARNING
WARNING
WARNING
5815148-01
INTERFERENCE. Other equipment may interfere with the
system, even if that other equipment complies with CISPR
emission requirements.
LOSS OF MONITORING. To avoid losing the Hub and sensor
connectivity and thus patient monitoring, do not use an RFID
reader operating at the frequency of 2.45 GHz close to any
part of the Hub and sensors.
LOSS OF MONITORING. Use of known RF sources, such as
cell/portable phones, RFID, electronic article surveillance (EAS)
systems, diathermy, or other radio frequency (RF) emitting
equipment near the system may cause unexpected or
adverse operation of this system. Consult qualified personnel
regarding system configuration.
LOSS OF MONITORING. Use of this equipment adjacent
to or stacked with other equipment should be avoided
because it could result in improper operation due to electrical
interference, or to increased equipment temperature due
to heat generated by the other equipment . If such use is
necessary, this equipment and the other equipment should be
observed to verify that they are operating normally. Consult
qualified personnel regarding device/system configuration.
MR UNSAFE. Do not use the system in an environment
containing an MRI magnet to avoid risks to the patient, user,
or devices.
UNEXPECTED SYSTEM BEHAVIOR. Use only approved
accessories listed in the product manual. Other accessories
may cause unexpected system behavior, damage the
equipment or system, result in increased emissions or
decreased immunity of the equipment or system, or interfere
with the measurement.
INACCURATE READINGS. After transport or storage of the
device outside the specified operating temperature range,
always allow the device to stabilize back to operating
temperature range before applying power to it. Using the
device outside the specified operating environment may
result in inaccurate results.
INCORRECT ALARM PROFILE. Make sure that you enter the
correct date of birth for the patient to ensure a correct alarm
profile. A wrong alarm profile may result in false or missing
alarms.
INCORRECT OR MIXED PATIENT DATA. Always make sure that
you have discharged the previous patient, removed the
patient and Hub assignment, and unpaired the sensor before
you start monitoring a new patient. Failure to do so will result
in mixed patient data.
Portrait
Mobile Monitoring Solution
Safety and intended use
25