Cleaning, disinfection, and care
Cleaning the alarming unit
In case there is some dried or gross soil, perform pre-cleaning by wiping with a soft,
clean non-linting cloth saturated with lukewarm tap water (temperature range 27 to
44°C, 81 to 111°F) until visible dried or gross soil is removed.
1. Dampen a soft non-linting cloth with detergent or use ready-to-use detergent
wipes.
Prepare the detergent according to the manufacturer's instructions.
2. Wring excess liquid from the cloth (wet but not dripping).
3. Wipe the exterior surface. Ensure that the entire surface is wet and make sure all
surfaces are uniformly cleaned.
Pay special attention to hard-to-clean areas, like grooves and crevices. Use
cotton swabs to access hard to reach areas. Do not use any sharp tools to
clean the device.
4. Replace soiled and/or dry wipes or non-linting cloths as needed.
5. Repeat the cleaning until the device is visibly clean.
6. Discard disposable wipes or non-linting cloths according to your hospital
guidelines. Do not reuse wipes.
7. Rinse the device thoroughly using a soft, clean, non-linting cloth saturated with
lukewarm tap water (temperature range 27 to 44°C, 81 to 111°F).
8. Dry thoroughly with a soft, clean, non-linting cloth before taking it into use.
Post-cleaning inspection
Inspect devices visually after cleaning and prior to disinfection for the following:
Cleanliness
If there is any visible soil present, repeat the cleaning procedure until no visible
soil remains.
Damage, including, but not limited to, corrosion (rust, pitting), discoloration,
excessive scratches, flaking, cracks, and wear
Missing or worn device labeling or marking
Do not use devices that are not functioning properly or are damaged, or have
unrecognizable markings, or missing or worn device labeling or marking.

Disinfection

Disinfection points to note

Always clean before disinfecting.
Always dilute cleaning and disinfectant agents according to their manufacturer's
instructions. Always consider your hospital guidelines as well.
Use only the permitted substances.
Your hospital guidelines permitting, all cleaning and disinfection activities can
be carried out at the bedside.
Before drying a surface after wiping it, wait the minimum time required according
to the substance manufacturer's instructions.