Alarm messages
Alarm message
Hub:
Learning
●
Hub:
RR high
●
RR low
●
Central Viewer:
RR high
●
RR low
●
Technical messages
Alarm message
Hub:
Check Resp patch
●
Check patch
●
Central Viewer:
Check Resp patch
●
Hub:
Resp battery low
●
Central Viewer:
Resp battery low
●
Hub:
Change Resp battery
●
Central Viewer:
Change Resp battery
●
Hub:
Clean Resp battery
●
interface
Clean battery
●
interface
Central Viewer:
Clean Resp sensor
●
interface
Hub:
No Resp battery
●
No battery
●
Central Viewer:
No Resp battery
●
138
Location
Possible explanations
data
The patient's breathing pattern
●
is being learned because the
measurement has just been
started or because of very large
artifacts in the signal.
al. area
Measurement values are equal
●
to or outside the alarm limits.
Location
Possible explanations
Hub:
The respiration patch is not
properly connected to the
al. area
●
patient or the sensor is not
data
●
properly connected to the
respiration patch.
Central
Viewer:
al. area
●
al. area
Sensor battery charge is low.
●
al. area
Sensor battery is empty.
●
Hub:
There is dirt in the sensor or
battery.
al. area
●
data
●
Central
Viewer:
al. area
●
Hub:
Battery has been removed from
the sensor.
al. area
●
data
●
Central
Viewer:
al. area
●
Portrait
™
Mobile Monitoring Solution
Suggested actions
Replace the respiration patch.
●
Wait until the message
●
disappears.
Check the patient.
●
Check electrode placement
●
and skin to electrode contact.
Adjust alarm limits on the
●
Central Viewer if necessary.
Suggested actions
Check the respiration patch
●
placement and skin to
electrode contact.
Check the sensor and
●
respiration patch connection.
Charge the battery.
●
Replace the battery.
●
Replace the battery and
●
put the empty one into the
charger.
Check and clean the sensor
●
and battery.
Connect a sensor battery.
●
5815148-01