GE Portrait Hub01 User Manual page 138

Mobile monitoring solution
Table of Contents

Advertisement

Alarm messages
Alarm message
Hub:
Learning
Hub:
RR high
RR low
Central Viewer:
RR high
RR low
Technical messages
Alarm message
Hub:
Check Resp patch
Check patch
Central Viewer:
Check Resp patch
Hub:
Resp battery low
Central Viewer:
Resp battery low
Hub:
Change Resp battery
Central Viewer:
Change Resp battery
Hub:
Clean Resp battery
interface
Clean battery
interface
Central Viewer:
Clean Resp sensor
interface
Hub:
No Resp battery
No battery
Central Viewer:
No Resp battery
138
Location
Possible explanations
data
The patient's breathing pattern
is being learned because the
measurement has just been
started or because of very large
artifacts in the signal.
al. area
Measurement values are equal
to or outside the alarm limits.
Location
Possible explanations
Hub:
The respiration patch is not
properly connected to the
al. area
patient or the sensor is not
data
properly connected to the
respiration patch.
Central
Viewer:
al. area
al. area
Sensor battery charge is low.
al. area
Sensor battery is empty.
Hub:
There is dirt in the sensor or
battery.
al. area
data
Central
Viewer:
al. area
Hub:
Battery has been removed from
the sensor.
al. area
data
Central
Viewer:
al. area
Portrait
Mobile Monitoring Solution
Suggested actions
Replace the respiration patch.
Wait until the message
disappears.
Check the patient.
Check electrode placement
and skin to electrode contact.
Adjust alarm limits on the
Central Viewer if necessary.
Suggested actions
Check the respiration patch
placement and skin to
electrode contact.
Check the sensor and
respiration patch connection.
Charge the battery.
Replace the battery.
Replace the battery and
put the empty one into the
charger.
Check and clean the sensor
and battery.
Connect a sensor battery.
5815148-01