Graphical user interface
2. Activate the screen. Note that you may need to enter the passcode to unlock
the Hub.
The LIVE view opens showing the numeric value(s) for the measured parameter(s).
The LIVE view can contain data for one or more parameters, depending on the
paired sensors.
3. If you cannot see the parameter on screen, make sure that the sensor has been
correctly paired.
If you can see the parameter data area for this measurement without any
●
values, the measurement is not connected to the patient, or there is a problem
with the patient connection. Check all connections.
You can pair sensors by selecting the plus indicator at the bottom left of the
screen:
Hub detailed view
The Hub detailed view shows the parameter name, alarm limits, numeric
measurement data, and cascaded waveforms of the measured parameter(s). Also the
signal quality indicators
displayed under the parameter label and unit of measure indication.
Viewing the Hub detailed view
1. Unlock the Hub screen if it is locked.
The LIVE view opens.
2. Touch the data area of the parameter you wish to see, or swipe your finger from
right to left.
The detailed view opens, showing the numeric values and waveforms of the
selected parameter.
60
are visible in the detailed view. The sensor battery status is
In this example:
1–11 = room-bed identification
●
Smith, Peter = patient's last name, first name
●
SpO2% (SpO2%) = Pulse oximetry label and unit of measure
●
90/OFF (90/OFF) = Pulse oximetry alarm limits
●
critical low: 87 (critical low: xx, where xx is 87 in this
●
example) = Critical low limit set for SpO
Numeric values 98 and 65 = SpO
●
values
PR/min (PR/min) = Pulse rate label and unit of measure
●
60/140 = Pulse rate alarm limits
●
in the waveform area = touch the arrow or swipe your
●
finger from left to right to return to the LIVE view
= Unpair the sensor
●
= Check connection of the selected parameter
●