11.EMC Tables
Guidance and manufacturer's declaration – electromagnetic emissions
The Fast-Fill is intended for use in the electromagnetic environment specified below. The
customer or the user of the Fast-Fill should assure that it is used in such an environment.
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC61000-3-2
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The Fast-Fill is intended for use in the electromagnetic environment specified below. The
customer or the user of the Fast-Fill should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Electrostatic
+/- 8 kV contact
discharge
(ESD) IEC
+/- 2 kV, +/- 4
61000-4-2
kV, +/- 8 kV, +/-
15 kV air
Electrical fast
±2kV
transients/burst
100kHz
s
repetition
IEC 61000-4-4
frequency
11 EMC Tables
Complian
Electromagnetic environment - guidance
ce
The Fast-Fill uses RF energy only for its
internal function. Therefore, its RF emissions
Group 1
are very low and are not likely to cause any
interference in nearby electronic equipment.
Class B
The Fast-Fill is suitable for use in all
establishments,
Class A
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
Complies
domestic purposes.
Compliance
level
+/- 8 kV contact
+/- 2 kV, +/- 4
kV, +/- 8 kV, +/-
15 kV air
±2kV
100kHz
repetition
frequency
31 / 37
including
domestic
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30 %.
Mains power quality should be
that of a typical commercial or
hospital environment.