solution manufacturer.
The water quality has to be convenient to the local regulations especially for the last
rinsing step or with a washer-disinfector.
Thoroughly clean and wash the components before autoclaving.
Do not use bleach or chloride disinfectant materials.
8.3.3 Autoclavable Components
Autoclavable Components
Cartridge
nut
Silver
tubes
WARNING
Only the components above can be autoclaved.
Before first use and after each use, sterilize the above components.
The Cartridge Nut, Silver Tube and Bending Tool must be cleaned, disinfected and
sterilized before first use following step 3 to step 9. After each use, please follow step
1 to step 9.
STE
INSTRUCTIONS
P
NO.
Initial
treatment
1
at point of
use
8 Maintenance
Immediately after using, wipe gross contaminations from
the components.
Prepare the components directly after treatment.
Do not submerge the components or wipe them with any of
the following functional water (acidic electrolyzed water,
23 / 37
Bending
Tool