Preparatio
n
before
2
cleaning
The following Step 3 to Step 5 are operated in a washer-disinfector:
WARNING
Use only approved washer-disinfectors according to EN ISO 15883, maintain
and calibrate it regularly.
Follow instructions and observe concentrations given by the manufacturer (see
general recommendations).
Sufficient rinsing step should be available in purified water (max 10 germs/ml
and max 0.25 endotoxin units/ml)
Avoid any contact between the cartridge nut and any instrument, kit, support or
container.
Make sure the components are dry before moving to the #6 step.
Cleaning:
Automated
3
8 Maintenance
strong alkaline solution, or ozone water), medical agents
(glutaral, etc.), or any other special types of water or
commercial cleaning liquids. Such liquids may result in
metal corrosion and adhesion of the residual medical
agents to the components.
Remove and disconnect the Cartridge from handpiece.
Refer to "Chapter 4-Installing the Fast-Fill" of this manual
for disassembly instructions. Then, cut out the Silver Tube
from Cartridge. Refer to "Chapter 8.2-Changing silver tube
of cartridge". Put them in container for transportation.
Observe suitable personal protective measures.
The Silver Tube is single use only and shall be
disposed of according to the local environmental
protection laws and regulation after use.
Carefully put the components (Cartridge nut, Bending Tool
and a new Sliver Tube) into the washer-disinfector and set
the parameters as follows:
-
Pre-clean with cold tap water for 2 min.
-
Clean with cleanser 55 ℃
demineralized water. Refer to the instructions for
24 / 37
for 5 min with

Advertisement