Troubleshooting
30. Gently press the tool on the slide
while releasing the button.
31. Lift the slide and remove it from
the scanner.
32. Remove the slide from the pick-up
tool.
33. Place the slide pick-up tool in the
maintenance door (I).
34. Check the removed slide.
35. If it is damaged or broken, make
sure to remove all glass particles
from the system.
WARNING
Make sure no objects are blocking the shutter.
36. Use the shutter knob to close the
shutter (I).
90
I
II
I
I
Pathology Scanner SG20 / SG60 / SG300
Troubleshooting procedures
III