Glass Slide Racks; Digital Slide Image - Philips Pathology Scanner SG20 Instructions For Use Manual

Table of Contents

Advertisement

Glass slide racks

Definition
Slide label on top of a cover slip
Cover slip on top of a slide label
Orientation
Tab. 42: Slide label specification
Glass slide racks
Definition
Type
Max. number of slides
Tab. 43: Glass slide racks
NOTICE
Only use clean glass slide racks.
Prevent accumulation of dust and wax particles inside the scanner as this can interfere with
the scanning process.

Digital slide image

Images of the scanner meet the following specifications:
Parameter
Pixel size (40x image)
Tab. 44: Pixel size
Parameter
Image compression level
Output format
Tab. 45: Image format
Parameter
Data format label image
Tab. 46: Label image
Pathology Scanner SG20 / SG60 / SG300
Specification
At least 18 mm x 18 mm of the cover slip is not covered by or not on
top of a label.
Max ± 5 degrees relative to glass slide edge
Specification
Winlab LS-20
Winlab LSM-20
Sakura 4768 20-Slide Basket
20
Specification
0.25 μm
Specification
The scanner supports 1 level of compression:
Level 2 = 15 x compression
iSyntax, Philips proprietary format
(for all image compression levels & RAW).
Specification
JPEG
Technical data