Troubleshooting
14. Gently press the tool on the slide
while releasing the button.
15. Lift the slide and remove it from
the scanner.
16. Remove the slide from the pick-up
tool.
17. Place the slide pick-up tool in the
maintenance door (I).
18. Check the removed slide.
19. If it is damaged or broken, make
sure to remove all glass particles
from the system.
Remove the slide from the stage area
20. Take the slide retrieval tool from
the maintenance door.
21. Open the slide retrieval tool
completely.
88
I
II
I
I
Pathology Scanner SG20 / SG60 / SG300
Troubleshooting procedures
III