Operation
Item to check
Make sure the barcode orientation is either vertical or horizontal.
Make sure the slide label is firmly affixed to the glass slide.
Avoid dirt, bleaching, serious staining and scratches on the barcode
area.
No additional markers added.
Make sure the slide label is applied on the upper side of the glass
slide.
Make sure the slide label is not rotated more than five degrees.
Make sure the slide label does not run on top of the cover slip.
Make sure the slide label does not stick out of the glass slide.
No oversized or undersized slide label.
No folded slide labels.
Not more than three slide labels on top of each other.
Make sure:
• only one barcode is added, when the barcode type and barcode
location are not configured.
• the correct barcode type is added and/or the barcode is added
on the correct location, when the barcode type and/or barcode
location are configured.
Tab. 10: Barcode checklist
32
Preparation of slides
Barcode printed on
Barcode printed on
label
slide
Not applicable
Not applicable
Not applicable
Not applicable