Installation of ISO-K Flange

For ISO-K flange connections, fix the two flanges with the clamps model
IC63250DCMZ or equivalent, as shown in the following figure (a protective screen
can optionally be used).
Figure 77
For models with TwisTorr 74 FS, TwisTorr 305 FS and TwisTorr 305 FSQ, use the
required number of claw clamps:
4 for ISO 63 flange
4 for ISO 100 flange
4 for ISO 160 flange
Tighten the bolts with a torque of 22 Nm.
For models with Turbo-V 551 use the required number of claw clamps:
10 for ISO 160 flange
Tighten the bolts with a torque of 25 Nm.