Emerson Rosemount Analytical FCLi Instruction Manual page 3

Free chlorine measuring system
Table of Contents

Advertisement

QUICK START GUIDE
1. Refer to Section 2.0 for installation instructions, and Section 3.0 for wiring instructions.
2. Once connections are secured and verified, apply power to the analyzer.
3. When the analyzer is powered up for the first time, Quick Start screens appear. Using Quick Start is easy.
a. A blinking field shows the position of the cursor.
b. Use the t or u key to move the cursor left or right. Use the p or q key to move the cursor up or down or to
increase or decrease the value of a digit. Use the p or q key to move the decimal point.
c. Press ENTER to store a setting. Press EXIT to leave without storing changes. Pressing EXIT also returns the
display to the previous screen.
English
Français
Español
# of sensors?
One
S1 Chlorine Type
free
total
S2 Measure?
Redox
Temperature in?
°C
FOR FCLi ANALYZER
4. Choose the desired language. Choose >> to show more choices.
>>
5. This screen appears if you have Model FCLi-02 (free chlorine and pH). Choose
two.
Two
If you have Model FCLi-01 (free chlorine only), the screen does not appear.
>>
6. Choose free for sensor 1.
pH
7. Choose pH for sensor 2. If you have Model FCLi-01 (free chlorine only), this
ORP
screen does not appear.
8. Choose temperature units.
°F
9. The main display appears. The outputs and alarms are assigned to default values.
10. Configure the analyzer for manual pH correction. Go to the main menu and
choose Program followed by Measurement. Choose Sensor 1 then Free
Chlorine. Choose Manual. In the next screen, enter pH 7,00. Do not choose
Auto.
11. To change outputs, alarms, and temperature-related settings, go to the main
menu and choose Program. Follow the prompts. For a guide to the Program
menu, see the menu tree on the following page.
12. To return the analyzer to the default settings, choose Reset Analyzer in the
Program menu.