Operation
Using ILV
On the Slave unit
Set the basic ventilation pattern:
As described for the Master unit, in accordance with
its respective Instructions for Use.
WARNING !
On the Slave ventilator, the rate setting »f«
does not have any direct effect during ILV.
However, as a safety precaution »f« should be
set to the same value as that ot the Master
ventilator. This will ensure that the two lung
compartments are not ventilated at different
rates in case the connection between the two
ventilators is broken.
Therefore:
Set »f« to the same value as on the Master unit.
Start ILV ventilation
Press and hold the key »Other Modes« on both
ventilators until the yellow indicator lights are
continuously lit.
ILV is now effective!
Select status page »ILV«.
Briefly press the screen key »Other Modes«
Screen display (example):
To return to the previous screen display:
Q
2 Press screen key »
Set Sigh (intermittent PEEP) following Evita 2 dura
Operating Instructions.