Dräger Evita 2 dura Ventilation Plus Addendum page 15

Table of Contents

Advertisement

On the Slave unit
If the second ventilator is an Evita 4 or Evita, refer to their
respective Operating Instructions for information on
settings.
If the second ventilator is an Evita 2 dura, set either:
Slave-synchronous I:E:
The I:E ratio of the Slave unit is synchronized directly.
Or:
Slave-asynchronous I:E:
The begin of inspiration and the total duration of an inspi-
ratory cycle are controlled by the Master ventilator;
the I:E ratio of the Slave unit can be selected as
required.
Or:
Slave-inverse I:E:
Inspiration of the Slave ventilator begins with expiration of
the Master unit and expiration of the Slave begins with
inspiration of the Master.
Do not activate ILV yet!
On the Master unit
Set the basic ventilation pattern:
Use the parameter keys and central dial knob:
Tidal volume »V
«
T
Respiratory rate »f«
Inspiratory time »T
«
insp
Inspiratory flow »Flow«
Positive endexpiratory airway pressure »PEEP«
O
concentration »O
vol.%«
2
2
Also: Pressure limitation »P
« if required.
max
Operation
Using ILV
15

Advertisement

Table of Contents
loading