Preventing Electrical Overstress Damage

Excessive RF power or voltage applied to the ports and connectors on the front
panel will damage the analyzer. Each port and front-panel connector has a
label (see photo below) that shows damage levels. These levels must not be
exceeded. Damage levels are also listed in the Specifications section of the
analyzer's Help system.
There are two ways to protect the PNA input ports from high RF power:
1. Use power limiters at the ports. Information on limiters can be found at:
2. Apply the recommendations in PNA Application Note 1408-10,
Installation Note E8356-90001
"Contacting Keysight" on page
www.keysight.com/find/mta.
"Recommendations for Testing High-Power Amplifiers," found at:
https://www.keysight.com/us/en/assets/9921-01766/user-manuals/User
s-Guide-A-Series-N991xA-2xA-3xA-5xA-6xA-Unabridged.pdf
(5989-1349EN).
4.
Keysight Part Number
9300-1367
9300-0980
9300-0797