Instruction Manual
90002929
07/2006
19.
Cleaning of Housing Outside
Use a liquid general purpose detergent and a lint-free cloth for cleaning the analyzer´s outside.
Procedure
Disconnect instrument from mains!
If opening the gas paths is required seal the open analyzer´s gas fittings
utilizing PVC caps to avoid pollution of inner gas path.
Moisten the lint-free cloth with a mixture of 3 parts of water and 1 part of the general purpose
detergent.
Do NOT drench the cloth, just moisten it to avoid liquid entering the
housing!
Clean the analyzer housing outside with the moistend cloth
If required dry the housing after cleaning
Emerson Process Management GmbH & Co. OHG
Before opening gas paths they must be purged with ambient air or
neutral gas (N
) to avoid hazards caused by toxic, flammable,
2
explosive or harmful to health sample gas components!
NGA 2000 MLT Hardware
CLEANING OF HOUSING OUTSIDE
19 - 1