CHECK AND REPLACEMENT OF ELECTROCHEMICAL OXYGEN SENSOR
BASIC CONDITIONS FOR EO
18.3.4
Basic conditions for the EO
Admit ambient air for the analyzer (approx. 21 Vol. - O
Connect the DVM to the measuring points
Tp 1 (Signal) and Tp 2 ( ⊥ ⊥ ⊥ ⊥ ⊥ ) of the PCB OXS, mounted directly at the sensor block
(Fig. 18-1, see also Fig. 18-2, 1-3, 1-16a and 1-17).
Set the signal to 1000 mV DC (± 5 mV) with potentiometer R4 (Fig. 18-3) of the
corresponding circuit board "OXS".
Note !
It is not allowed to change this setting for this sensor again !
Switch off the analyzer and close the analyzer housing (see 15.).
Built-in the module into platform if necessary.
A complete re-calibration of the instrument must be performed after a sensor - replacement.
18 - 6
-SENSOR
2
Tp 1
Potentiometer "R4"
Fig. 18-3: PCB "OXS", assembled, horizontal projection
-Sensor
2
) and switch on (see Section 6.).
2
Tp 2
Connection
oxygen sensor
("P2")