NGA 2000 MLT Hardware
TROUBLESHOOTING
DISPLAY MESSAGES
10.3.3

Detector signal communication failed

Possible Causes
a)
Failure in signal processing:
Detector failure
b)
Contamination of the gas paths
(
10.3.4

Light source

Possible Causes
a)
Failure in signal processing:
Light source failure
10.3.5

Detector

Possible Causes
a)
Failure in signal processing:
Detector failure
b)
Contamination of the gas paths
10 - 4
Check / Correct
Check connection to PCB "PIC" Fig. 1-32) or
PCB "DSP" Fig. 1-33) resp.;
See Section 11.1.4 (IR/UV measurement),
Section 11.1.6 (O
measurement).
2
Check if photometric components are dirty,
clean or replace the components if necessary
(cf. Section 17.).
Check and adjust physical zero point alignment
see Section 17.5 / 11.1.6)
Check / Correct
Check connection to PCB "PIC" Fig. 1-32) or
PCB "DSP" Fig. 1-33) resp.;
See Section 11.1.2
Check / Correct
Check connection to PCB "PIC" Fig. 1-32) or
PCB "DSP" Fig. 1-33) resp.;
See Section 11.1.4
Check if photometric components are dirty,
clean or replace the components if necessary
(cf. Section 17.).
Check and adjust physical zero point alignment
(see Section 11.1.6)