MLT 2 (Field Housing)

The MLT 2 variation is designed for wall mounting and is available in two different versions: One
version, equipped with standard front panel is intended to be installed at protected locations
whereas the version with impact tested magnetically operated front panel may be outdoor
installed.
In any case the restrictions for ambient temperatures have to be taken into account, see section
technical data!
This current instruction manual covers using the MLT 2 analyzer for general
purpose applications only!
Installation, startup and maintenance for operation in hazardous areas are
described in detail in a separate instruction manual, shipped together with
each such analyzer and are not subject of the current instruction manual!
5.4.2.1

Wall Mounting

This housing is designed for wall mounting.
Utilizing the four mounting brackets at the analyzer's rear side the instrument can be wall mounted.
For fastening points see Fig. 5-6 please.
The analyzer variations MLT 2 intended to be wall mounted and/or outdoor
installed weigh up to 35 kg, depending on included options!
Use two person and/or suitable tools for transportation and lifting these
instruments!
Take care to use anchors and bolts specified to be used for the weight of
the units!
Take care the wall or stand the unit is intended to be installed at is solid
and stable to hold the units!
Emerson Process Management GmbH & Co. OHG
MLT 2 -- HEAVY INSTRUMENTS !
NGA 2000 MLT Hardware
ANALYZER SPECIFIC INSTRUCTIONS: MLT 2
INSTALLATION
5 - 13