Operation

Warning
If the unit is tilted in excess of 30
degrees, do not apply electrical
power for a minimum of 12 hours.
Warning
This product is NOT approved for
storage of flammable or explosive
materials. Also, it is NOT approved
for use in hazardous locations con-
taining explosive atmospheres.
Note
The refrigerator has a built in delay
time before compressor activations.
Upon initial start-up, the unit will delay
before cooling begins.
Figure 3
8
Before operation, become familiar with the refrigerator
controls located on the refrigerator header panel. A layout
of the controls is given below.
Begin operation by inserting the key into the key switch
located on the header panel. The key switch is packed
inside the envelope, which is shipped in the refrigerator
chamber. Turning the key switch to the ON ( | ) position
will energize the evaporator fans and the digital controller.

Controls Layout

The digital temperature controller is located on the header
panel (See Figure 3). When the unit is initially turned on,
the display will indicate current chamber temperature.
The temperature units will be indicated by the °C or °F
LED located just to the right of the temperature display.
The refrigerator is factory set at 4°C. When the unit is first
energized the evaporator fans will operate, however it will
not immediately cool. A compressor delay time is pro-
grammed into the controller to verify the system pressure
is low enough to activate the motor. When the delay time
is over, the Cool LED will illuminate, the compressor will
run and the chamber will cool.
I
ON
O
OFF
Mode LED
Menu Keypad
Increase Keypad
Decrease Keypad
Interior Light Switch
Temperature Display
Units LED'S
POWER
O
Cool LED
Key Switch
I